The Advantage™ Ultrasound Phantom is an easy to use comprehensive means of buating ultrasound systems with an effective operating frequency range of 2.25 to 10MHz. The phantom is designed with a combination of monofilament line targets for distance measurements and tissue mimicking target structures of varying sizes and contrast content. Four gray scale targets ranging in contrast from +6 to -3dB are provided to buate the system’s displayed dynamic range and gray scale processing perbance.
The Advantage™ Phantom offers a new and improved dual scan surface design to easily accommodate linear, sector, endoscopic probes and transesophageal (TEE) probes.
scanning-port-for-tee-probes-phantom-picenoscopic-scan-surface-with-trough-phantom-picstandard-scan-surface-phantom-pic
Dead Zone or Ring-Down (A1 & A2)
Horizontal Measurement Calibration Linear (C1&C2)
Horizontal Measurement Calibration Sector (C1 & C2)
Sensitivity/Penetration (F)
Vertical Measurement Calibration (B1 & B2)
Image Unibity (target-free areas)
Functional Resolution (F1, F2)
Focal Zone (B1, B2 or F)
Axial & Lateral Resolution (E1, E2, E3)